BioWaveHOME

There are 3 buttons that control the device:

1. Power ON/OFF button

2. PLUS (+) button to increase intensity

3. MINUS (–) button to decrease intensity

4. PAUSE button to pause treatment 

IMPORTANT:

  • Turn on device AFTER you have completed the below
    set up steps on the next page.
  • Please charge battery at least 1x per week.

BioWaveHOME - Directions for Use

Do not turn on unit until Step 6

1.

Clean your skin with a damp washcloth in the location the electrodes are to be placed.

2.

Plug ELECTRODE CONNECTORS into blue LEADWIRE CABLE CONNECTORS. Either electrode can be attached to either blue leadwire cable.

3.

Place electrodes on body. At least one ROUND electrode must be placed over the pain site. See and follow Electrode Placement Examples inside.

4.

Orient LEVER on cable CONNECTOR so it is aligned with its mating keyhole inside the CABLE RECEPTOR in the device. GENTLY slide cable CONNECTOR into CABLE RECEPTACLE so it clicks in place.

5.

Press Power Button to turn unit on. Start up screen should read 0.0%.

6.

Start treatment by pressing the PLUS (+) button.

7.

Continue to press the PLUS (+) button throughout the treatment so that a steady strong but comfortable tingling and pressure sensation is felt under the electrode(s) covering the pain site(s).

8.

Unit turns off automatically at end of 30 minute treatment.

9.

Plug AC Charger into AC RECEPTACLE and other end into a standard wall outlet to recharge the battery. Unit provides up to 4 treatments on a single charge.

See User’s Manual for detailed instructions. 

Getting started videos:

Electrode Placement Rationale

BioWaveHOME electrode placement are different from conventional electrical stimulation. The mixing of the patented therapeutic signals occurs in a 3.5 inch hemisphere (the size of half of a grapefruit) beneath and surrounding each electrode, not along the surface of the skin between the electrodes. As a result, electrodes need to be placed directly over specific locations of pain, or one electrode over a single location to receive stimulation – typically over a bony prominence near the treatment site. 

Electrical signals can be focused to different parts of the body by pairing electrodes of different sizes with one another. 

AREAS OF TREATMENT

B-Set: Two Locations of Of Pain or one larger area of pain. Two same size 2″ diameter round Pain Site Electrodes.

E-Set: Single Location of Pain on Extremities. One 1.375″ diameter round Pain Site Electrode; One 2″ x 4″ rectangular Dispersive Electrode placed in a comfortable location (over a bony prominence).

U-Set: Single Location of Pain in the Mid-Torso Area. One 2″ diameter round Pain Site Electrode; One 5″ x 8″ large rectangular Dispersive Electrode placed in a comfortable location (horizontally across the back).

Pads must not touch each other when the unit is turned on – one inch is the minimum spacing.

Pads are reusable but must be tacky to the touch. If not, please use a new set of pads.

Pads can typically be reused 8-10 times.

B-Set: Two Locations of Of Pain or one larger area of pain. Two same size 2″ diameter round Pain Site Electrodes.

E-Set: Single Location of Pain on Extremities. One 1.375″ diameter round Pain Site Electrode; One 2″ x 4″ rectangular Dispersive Electrode placed in a comfortable location (over a bony prominence).

U-Set: Single Location of Pain in the Mid-Torso Area. One 2″ diameter round Pain Site Electrode; One 5″ x 8″ large rectangular Dispersive Electrode placed in a comfortable location (horizontally across the back).

Pads must not touch each other when the unit is turned on – one inch is the minimum spacing.

Pads are reusable but must be tacky to the touch. If not, please use a new set of pads.

Pads can typically be reused 8-10 times.

AREAS OF TREATMENT

NECK

BioWave On Neck Demonstration

Neck or cervical pain in two locations

BioWave On Neck Demonstration

Bilateral neck or cervical pain

BioWave On Neck Demonstration

Neck or cervical pain in
multiple discs

SHOULDER

BioWave On Shoulder Demonstration

Shoulder pain in two locations or Frozen Shoulder

BioWave On Shoulder Demonstration

Pain at top or throughout shoulder (e.g. AC Sprain or Frozen Shoulder)

BioWave On Shoulder Demonstration

Pain in edge of shoulder
(e.g. Rotator Cuff Tendinitis)

BioWave On Shoulder Demonstration

Pain in front of shoulder
(e.g. Biceps Tendinitis)

BioWave On Shoulder Demonstration

Pain in back of shoulder
(e.g. Infraspinatus Strain)

BioWave On Shoulder Demonstration

Trapezius pain in one
location (e.g. Trigger Point)

BioWave On Shoulder Demonstration

Trap pain in 2 locations
(e.g. Bilateral Trapezius pain)

BACK / TORSO / BUTTOCK

BioWave On Lower Back Demonstration

Back pain in 2 locations
(e.g. Bilateral low back pain)

BioWave On Lower Back Demonstration

Radiating back pain – one pad over origin of pain, one pad over first pain site (pads must be on skin)

BioWave On Lower Back Demonstration

Sacroiliac (SI) Joint Pain; pain from Interstitial Cystitis

BioWave On Lower Back Demonstration

Back pain over large area
(e.g. Rotational Strain)

BioWave On Lower Back Demonstration

Back pain in 1 location
(e.g. Unilateral low back pain
to the right of the spine)

BioWave On Lower Back Demonstration

Low back pain focused over a facet joint

BioWave On Lower Back Demonstration

Low back pain focused over
the spine

BioWave On Back Demonstration

Mid-back pain (e.g. Bilateral thoracic pain)

ELBOW / BICEPS / TRICEPS / FOREARM

BioWave On Elbow Demonstration

Pain in outside of elbow
(e.g. Tennis Elbow)

BioWave On Elbow Demonstration

Pain in inside of elbow
(e.g. Golf Elbow)

BioWave On Elbow Demonstration

Pain behind elbow
(e.g. Triceps Tendinitis)

WRIST / HANDS / FINGERS

BioWave On Wrist Demonstration

Pain in back of wrist. Second pad on front of wrist (e.g. Sprains, Strains, Tendinosis)

BioWave On Wrist Demonstration

Pain in front of wrist. Second
pad on back of wrist (e.g. Sprains, Strains, Tendinosis)

BioWave On Hand Demonstration

Thumb pain (e.g. thumb or
UCL Sprain)

BioWave On Hand Demonstration

Pain in finger joints

KNEE

BioWave On Knee Demonstration

Pain throughout entire knee
(e.g. OA or total knee
replacement)

BioWave On Knee Demonstration

Central knee pain near front of knee (e.g. ACL Sprain)

BioWave On Knee Demonstration

Pain below kneecap
(e.g. Patellar Tendinitis)

BioWave On Knee Demonstration

Pain in inside of knee
(e.g. Bursitis, OA, MCL Sprain)

BioWave On Knee Demonstration

Pain in outside of knee
(e.g. Bursitis, OA, LCL Sprain)

BioWave On Knee Demonstration

Pain in Illiotibial (IT) band

BioWave On Knee Demonstration

Pain above knee
(e.g. Quadriceps Tendinitis)

BioWave On Knee Demonstration

Pain in back side of knee
(e.g. PCL Sprain)

HIPS

BioWave On Lower Back Demonstration

Hip pain

quadriceps

BioWave On Upper Leg Demonstration

Quadriceps pain

hamstring

BioWave On Lower Leg Demonstration

Hamstring pain

calf

BioWave On Lower Leg Demonstration

Calf pain – one pad over pain site (e.g. Gastrocnemius Strain) other pad is placed toward bottom of knee Cap

SHIN

BioWave On Lower Leg Demonstration

Shin pain (e.g. Shin Splints)

ankle / feet

BioWave On Ankle Demonstration

Pain from Diabetic Neuropathy (Tibial Nerve Stimulation)

BioWave On Ankle Demonstration

Ankle or foot pain on outside of foot (e.g. Sprain)

BioWave On Ankle Demonstration

Ankle or Foot Pain on Inside
of Foot

BioWave On Foot Demonstration

Plantar Fasciitis

BioWave On Foot Demonstration

Achilles Tendinitis

BioWave On Foot Demonstration

Pain in top of foot. Second pad on bottom of foot under pad on top (e.g. Metatarsal or
Neuroma pain)

IMPORTANT

Contraindications
When Not to Use BioWaveHOME

  • Do NOT use BioWaveHOME if you have an implanted cardiac pacemaker.
  • Do NOT use BioWaveHOME if you have epilepsy or are prone to siezures.
  • Do NOT place electrodes over the heart, or across the thoracic volume.
    Electrodes can be applied to the back of the thorax and lateral aspect of
    the upper limb (i.e. below/down the shoulder).
  • Do NOT place electrodes on the front or side of the neck.
  • Do NOT place electrodes on top of the head.
  • Do NOT place electrodes over open wounds, broken skin, rashy skin or
    sensitive areas.

NOTE:
Electrodes MAY BE PLACED directly over or in the proximity of implanted metal hardware including total joint replacements, rods, plates, screws, anchors and pins. See User’s Manual for more information.

BioWave Pads Illustration

ELECTRODES (PADS) MUST NEVER TOUCH EACH OTHER DURING THE TREATMENT

  • 1.0 inch (2.6 cm) is the minimum spacing between Pads.
  • There is NO maximum spacing between any Pads.
  • If the edges of the Pads touch during the treatment,
    it may cause a burn.
  • Pads typically can be reused about 8-10 times.